Technique item |
Technique specifications: |
Overall maximum size: |
303.5*138*93mm |
Maximum weight: |
2.0 KG |
Power supply: |
AC100-240V.50/60HZ |
Max power consumption: |
25VA |
Applicable syringe: |
All kinds of standard syringes: 10ml, 20ml and 50ml |
Injection accuracy: |
±3%( after being adjusted, up to ±2%) |
Bolus rate: |
10ml: 100ml/h 20ml: 200ml/h 50ml: 500ml/h |
Displayed information: |
①injection rate ②accumulated injection volume ③syringe size④battery capacity ⑤battery charging indication��?AC power supply connection indication ⑦BED NO.��?alarm information |
Injection increment: |
0.1ml/h |
Preset injection range setting: |
0.1~999.9ml |
Accumulated injection volume display: |
0.1~999.9ml |
Battery: |
Rechargeable lithium polymer battery 7.4V, 1600mA Working for 4 hours (25ml/h) after fully being charged |
Electromagnetic compatibility: |
Meet the requirements of IEC60601-1-2(2001 version) |
Electrical safety: |
Meet the requirements of IEC60601-1:1995, GB9706.1-2007 |
Injection rate range setting: |
10ml: 0.1~100ml/h 20ml: 0.1~200ml/h 50ml: 0.1~500ml/h |
Alarm content: |