Model BD-8000 Intelligent Infusion Pump -Manufacturer
Highlights:
1) User-friendly Anti-slip design; 2) Big color LCD screen display; 3) Three work modes selection: Rate mode, body weight mode and time mode; 4) Intelligent factory default; 5) Adjustable alarm sound from 0-3 levels; 6) Adjustable air bubble and pressure value for alarm; 7) Adjustable KVO rate: 1-5ml/h; 8) Port of car power supply etc.
Unique Functions: 1) Unique door free structure avoids pump stuck and fluid leakage while Big LCD screen and English interface enabling easy operation. 2) Applicable for all brands of standard infusion set which meet the international standards, pressing the button for selection to ensure the infusion accurate, smooth and safe. 3) Audio-Visual alarm gives out and stop working automatically when the air detector detects there is air bubble in the pipe, in this way it can reduce the risk of causing air embolism. 4) Audio-Visual alarm gives out when the pressure sensor detects the pressure is beyond setting range. 5) Built-in double CPU monitor to ensure the infusion process safer and more reliable with wide infusion rate range from 1~2000ml/h. 6) With bolus (fast speed) dosage function: 2000ml/h.
Specification: |
|
Product Model |
Model: BD-8000 Infusion Pump |
Infusion Pump Mechanism |
Peristaltic Mechanism |
Flow Rate Range |
1-2000 ml/h |
Bolus Rate |
1000 ml/h |
Infusion Increment |
0.1 |
Water-proof Classification |
IP21 |
Infusion Mode |
1. Rate Mode; 2. Time Mode; 3. Body Weight Mode; |
Rate Mode |
Rate: 1-2000 ml/h |
Time Mode |
1. Time: 1minute-6000minutes 2. Vol.: 0.1ml-9999.9 ml |
Body Weight Mode |
1. Body Weight: 0.1kg-300.0 kg 2. Drug: 0.1mg-999.9 mg 3. Liquid: 0.1-999.9 ml 4. Dose: 0.1-9999.9 (The maximum value is restricted to body weight, drug, liquid, dose and unit) 5. Dose Unit: mg/kg/h or ug/kg/min |
KVO Rate |
Adjustable Range: 1ml/h-5.0 ml/h (The pump will infuse at the KVO rate when there is an occlusion alarm) |
Preset Volume Range |
0.1-9999.9 ml |
Preset Time Range |
0-99 hours 59 minutes |
Accumulated Infusion Volume Displayed |
0.1-9999.9 ml |
Infusion Accuracy |
± 5% |
Power Supply |
AC100V-240V, allowable error± 10%, 50/60Hz or D. C. 7.4v, relative error +5%, -10% (rechargeable lithium battery, 1600mAh) |
Maximum Power Consumption |
Max25VA, running more than 2 hours at the rate of 25ml/h after being fully recharged. |
Battery Charge |
When infusion pump is connected to AC power, turn on the Power and the battery will start charging automatically (about 8-14 hours to charge fully). |
Fuses |
Double 250V 2A |
Displayed Information |
Flow rate, preset volume, accumulated volume, Battery capacity, AC power symbol, infusion completion, occlusion, air bubbles, bed No. and pressure value etc. |
Status Indication Information |
Stop, Run, Bolus, KVO |
Alarm Information |
Over, occlusion, air bubbles, low pressure, installation error of infusion pipe, abnormal1, abnormal2, drops error and serious low battery. |
Air Bubble Detection |
At the rate of 600ml/h, the smallest bubbles that can be detected are 0.005ml of volume. |
Infusion Pressure |
Maximum pressure is 300kPa, occlusion gate value is 39.2kPa-137.2kPa. The longest time to give occlusion alarm is 2 minutes at rate of 25ml/h. And at rate of 5ml/h is 5 minutes. |
Maximum Size of Outer Shell |
125mm× 140mm× 205mm (length× width× height) |
Maximum Weight |
< 3.0 kg |
Classification |
Class II, Type BF |
Work Mode |
Continuous Operation |
Outer Shell Material |
ABS Plastic |
Operating Conditions |
Environment temperature: 5degrees celsius-+40degrees celsius; Air pressure: (86-106)Kpa; Relative humidity: 20%~80% |
Storage Conditions |
Environment temperature: -20degrees celsius-+50degrees celsius; Air pressure: (50-106)Kpa; Relative humidity ≤ 95%; Non-corrosive gas and well-ventilated room. |
Applicable Infusion Pipes |
National standardinfusion pipes |
Applied Standard |
GB/T191-2008 GB8368-2005 GB9706.1-2007 GB 9706.27-2005 GB9969.1 GB/T14710-1993 YY0466-2003 |
|